Dialog+ · Degassing Chamber Integrity Test (Pneumothorax Check)

The Pneumothorax Check — Structural Integrity of the Degassing Chamber

Pressure Leakage/Integrity Test on the degassing chamber loop should be done before AEP pump replacement.

In our medical analogy, this is the equivalent of checking for a pneumothorax—a structural leak in the wall of the lung that destroys the internal negative pressure vacuum.

Critical: If the chamber or its connecting seals leak even a microscopic amount of ambient air, the AEP pump will never be able to pull its mandatory -600 mmHg vacuum, causing the machine to continuously throw air-lock and pressure alarms.

1. Anatomy of the Leak Points (Where the "Lung" Fails)

Inspection Zones

Image Placeholder: Degassing Chamber Assembly — Leak Points

Insert photo: Labeled degassing chamber showing acrylic body, top vent port O-ring, PDA sensor interface, and inlet/outlet connections.

The degassing chamber block assembly on the B. Braun Dialog+ contains several material transition zones that are susceptible to fatigue:

Clinical Pearl: The acrylic body is the most common failure point. Thermal cycling causes microscopic cracks that are invisible to the naked eye but will leak vacuum under negative pressure.

2. Step-by-Step Technical Integrity Test (The Bench Protocol)

Procedure

Required Tools

[Handheld Mityvac Pump] │ (Apply +1.0 Bar Pressure)

[Chamber Top Port]

[Plugged] ─── [Degassing Chamber] ─── [Plugged]
(Bottom Inlet) (Outlet to Pump)

The Isolation & Testing Procedure

  1. Isolate Power and Fluid: Power down the Dialog+ completely and unplug it from the AC outlet. Turn off the external RO water supply line.
  2. Seal the Boundaries:
    • Disconnect the fluid input line at the bottom of the degassing chamber and plug it tightly with a luer-lock cap or clamp the silicone tube shut using hemostats.
    • Disconnect the fluid output line that runs from the chamber to the suction side of the AEP pump and plug/clamp it securely.
  3. Connect the Test Rig: Disconnect the top air vent line from the chamber. Connect the flexible tube of your manual handheld pressure pump directly to this upper chamber port.
  4. Pressurize the Organ (Positive Pressure Test):
    • Slowly pump the handheld test rig to inject air into the degassing chamber until the mechanical gauge reads exactly +1.0 bar (approx. 14.5 PSI) of positive pressure.
The Stopwatch Rule: 60-Second Hold Test

Stop pumping and start a timer for 60 seconds.

✅ PASS: Drops < 0.05 bar — Chamber is structurally sound
❌ FAIL: Drops > 0.05 bar — Active pneumatic leak confirmed

3. Pathological Sign Isolation (Finding the Crack)

Localization

If the pressure needle is actively dropping, teach your technicians how to physically localize the structural hole:

Image Placeholder: Bubble Solution Test in Progress

Insert photo: Soapy water applied to degassing chamber with bubbles forming at a hairline crack.

Clinical Reasoning: The bubble test is your definitive localization tool. It takes 2 minutes and prevents the costly mistake of replacing a perfectly good pump head when the real problem is a $5 O-ring or a cracked chamber.

4. Differential Diagnosis (Ruling out Mimics)

Rule Out

When performing the positive pressure test, the following can mimic a chamber leak:

5. Technical Management (The Repair Plan)

Treatment Plan
Temporary "Field" Fixes are Strictly Prohibited:

Instruct your staff that attempting to seal a cracked degassing chamber with epoxy, glue, or silicone is an unsafe practice.
The extreme vacuum (-600 mmHg) and high heat disinfection cycles will cause temporary sealants to fail mid-therapy, risking an emergency patient shutdown on the clinic floor.

The Action Plan

1
Total Organ Replacement (Cracked Body) If the acrylic body is cracked (confirmed by bubble test), the technician must replace the entire chamber column with a new B. Braun OEM Degassing Chamber Block Assembly.
Do not attempt to patch, glue, or repair the acrylic — replace it.
2
O-Ring Replacement (Seal Failure) If the leak is isolated to the top vent port or sensor port:
  1. Unscrew the fitting.
  2. Remove the flattened elastomer ring.
  3. Lubricate a brand-new EPDM chemical-resistant O-ring with a light film of medical-grade silicone grease.
  4. Re-torque the fitting securely (do not over-tighten).
3
Post-Op Verification (Mandatory) Always repeat the 60-second mechanical pressure test after reassembly to guarantee a 100% airtight loop before putting the machine back into service mode for system calibration.
Confirm: Pressure holds at +1.0 bar with < 0.05 bar drop over 60 seconds.
Final Sign-Off: Document the test results in the machine's service log. A failed test must be resolved before any AEP pump replacement is performed.
✍️ Author: Ahmed Mohmad Rashyd Musleh Registered Staff Nurse